POWER SUPPLY REDIRECTION TO AUXILIARY ADAPTER
6–5–1 Introduction
This section describes the procedures for replacing the power supply unit of LOGIQ 500.
6–5–2 Replacement Procedure
READ and UNDERSTAND these instructions thoroughly before proceeding with the replacement. Perform each step
in sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and
check for implementation errors before contacting your Ultrasound Region Engineer.
CAUTION
Possible operational damage. Failure to strictly follow ESD (Electrostatic Discharge) precautions during this upgrade may cause constant or intermittent operational abnormalities.
Strictly follow all precautions.
Check each step as it is completed.
Parts Required
FRU 717 |
PART NAME Power Supply AC110V |
QTY 1 |
PART # P9523DB |
REMARKS for 110–120V model |
718 Power Supply AC220V 1 P9523DA for 220–240V model
Time Required
Approximately four (4) hours
Tools Required
Standard field Service Tool kit
Three Jacks
Pallet (Height: 70mm, Width and Depth: suitable length)